Coil fields & geometry
Free-space coil fields and ideal RF excitation.
ROI metric: max | |B(r)| − |B(0)| | / |B(0)| × 100%, sampled over the circular x–z region. This measures magnitude variation in a 2D section, not 3D or B₁⁺ uniformity.
Field map · y = 0
Orange dots: wire crossings; orange dashes: coil planes (not conductors). White circle: ROI. Arrows show direction only. Color saturates above 3× the center field; the scale updates with current. Points within 0.03R of ideal wires are masked; this is not a finite-wire model.
Coil geometry
Signed field along the coil axis
Theory & experiments
Double the current: field strength doubles. Reverse it: arrows and the signed profile reverse. Double R at fixed I and N: center field halves. Select a pair and compare d/R = 0.5, 1, and 1.5; at d/R = 1 the second axial derivative vanishes at the midpoint.
The numerical solver sums 256 current elements per loop. N turns are treated as coincident loops. Near-wire fields are excluded because ideal filament fields diverge. Current is prescribed in amperes. The RF excitation tab interprets it as a peak sinusoidal amplitude.
RF field to excitation
Low-field context (<0.5 T): the carrier frequency is lower. Tuning, conductor/sample noise, matching and environmental interference remain future models.
Changing B₀ magnitude changes the on-resonance carrier frequency, not this prescribed-current B₁⁺ map or its flip angles. No detuning, frequency-dependent current, or tissue loading is calculated.
Both color scales are fixed. Values outside the locked ranges saturate at the colorbar ends. Unlock to adjust contrast.
At fixed coil current and geometry, changing pulse duration leaves the B₁⁺ field unchanged but scales the flip angle.
Window / level
Window / level
Both maps use the y = 0 section and the same near-wire exclusions as the coil map. White circle: ROI; dots: wire crossings. Window and level affect colors only, not field values or rotation angles. Values outside the displayed range saturate at the ends of the color scale. Full range includes near-wire peaks.
Why the factor of two?
A linear RF oscillation decomposes into two equal, oppositely rotating components. The co-rotating component drives on-resonance excitation in the rotating-wave approximation. A negative current reverses RF phase by 180° but leaves these magnitude maps unchanged. This half-amplitude rule is specific to the real, linearly polarized field modeled here.
This is an ideal hard-pulse model: uniform B₀, on resonance, no relaxation, no gradients, no RF phase map, and no counter-rotating or longitudinal RF dynamics. Local transverse magnetization magnitude is not a reconstructed image or a receive-coil signal.
RF polarization and rotating frames — Shaihan Malik · Proton gyromagnetic ratio — NIST
Next models · planned, not simulated
- Advanced excitation: complex RF phase, quadrature coils, off-resonance Bloch dynamics and shaped pulses (see the Bloch simulator for off-resonance dynamics).
- Tissue loading: conductivity and permittivity, sample losses and coil impedance.
- High-frequency waves: spatial phase, dielectric effects and full-wave field solutions.
- Low-field MRI (<0.5 T): tuning at lower Larmor frequencies, coil versus sample noise, conductor losses, matching and environmental interference. Low B₀ alone will not be used as a guarantee that loading or wave effects can be ignored.
The RF tab uses B₀ to set orientation and carrier frequency. Coil fields still depend only on geometry and prescribed current.